Bench-Top EC6S-MD

The bench-top EC6S-MD has the latest technology and integrated safety as a medical device of class IIa (93/42 / EEC Annex V).

  • Heated lid and bottom in every chamber
  • Peekaboo Windows for no disturbance
  • Tri-gas (CO2, O2 and ambient air) system
  • 6 separated and independent chambers
  • Fast temperature and gas recovery
  • Pre-heated gas mixture in built-in gas tank
  • Long life CO2 and O2 sensors
  • Analog output and BMS (contact relay) standard

Product facts and notices

External dimensions (mm): W517.5 x D608 x H171


Chamber dimensions (mm): W108 x D180 x H18.5


Chamber capacity: 300ml (with bottom plate attached)


Temperature control method: Digital PID control


Temperature control range: Room temperature +5ºC to 40ºC


Temperature fluctuation range: ±0.1ºC


CO2 sensor: Infrared type


CO2 concentration control range: 0% to 20.0%


CO2 concentration control method: Duty control


CO2 concentration fluctuation range: ±0.3%


O2 sensor: Ceramic oxygen sensor


O2 concentration control range: 0.1% to 18%


O2 concentration control method: Duty control


O2 concentration fluctuation range: ±0.5%


Safety device: Temperature increase prevention function (Software)


External output:

  • Analog output: Temperature,CO2/O2 concentration,gas flow rate
  • Contact output: Close when an alarm is sounded/ Close when power blackout occurs

Option: Can be connected to a PC incubator monitor system/ Monitoring with dedicated EZC-Logger software*2


Weight: 26Kg


Power consumption: 150 VA (max.)


Classifications:

  • EN 60601-1:2006/A1:2013/IEC60601-1:2005/A1:2012
  • EN 60601-1-2:2015/IEC 60601-1-2:2014
  • EN 62366:2008/IEC62366:200
  • EN 2304:2006/IEC62304:2006) 93/42/Eec (MDD)
         

Media

Downloads

Brochure Bench-Top EC6S-MD (3 MB)

EC Certificate (157 KB)